  • Nogicom

    CALIBRACIÓN BASCULA OHAUS AVIATOR 2000

    Short circuit cal jumper on pcb. Scale will show required weight. Place weight and press enter. F means finished. Automaticall exit. If you want another weight, change with C before pressing enter.

    hace 6 años
  • Nogicom

    Tru Test 703B "Service" on display

    I have "Service" on display after everything. I am using "700 Series Calibration and Configuration Manual.pdf", and I can calibrate it with simulator, or with real cells. But after succesfull cal it shows "trutest, no-trade, zeroing" sequence and then Service....

    0
    hace 8 años
  • Nogicom

    manual tecnico ind 570, metler toledo

    Instead of one TM there is UG + IM. TM only for Fill application. UG is 20 mb. Too much for upload. But you can download it here: https://www.mt.com/dam/product_organizations/industry/IndustrialTerminals/30205308_R07_MAN_IND570_UG_EN.pdf https://www.mt.com/dam/product_organizations/industry/IndustrialTerminals/30205319_R06_MAN_IND570_IM_ML.pdf

    hace 8 años
  • Nogicom

    METTLER TOLEDO MODELO PB 3002

    Beautiful balance. Electrical problems for quick check: 1.) Measure resistance of the coil - but beware, wires are hardly to see. Two same windings with the center tap. Resistance few hundred ohms. 1.a) Coil goes to the small 4pin connector...

    hace 8 años
  • Nogicom

    Balanza PB303 Mettel Toledo.

    See my comment on PB3002 up there.

    hace 8 años
  • Nogicom

    Necesito ajustar digi di-160

    Move switch on pcb UP to enable cal. Switch on scale. Scale must be empty. While 88888 on LCD for the second time, press and hold ZERO, then arrow LEFT and then 2 times TARA. Release ZERO. First is division...

    hace 9 años