hace 5 años
In this case, the fault is minor and clear . At the input of the multimeters there is a voltage resistor divider (in series), the resistance between the 20 and 200 scale is cut; as it is cut, mark more in 200 and nothing in 20 and lower. This divider makes 10 MOhm in total, and the resistances are 9M, 900K, 90K, 9K, 900 ..., you have to check them, one of them is cut.