⚡ Descuentos en cursos en video - Lleva tu aprendizaje técnico al siguiente nivel y aprovecha precios especiales antes de que termine la oferta. 🔥 Aprovechar oferta ahora 🔥
X-rays

radioscopy team siemens tridoros

walservi40
hace 5 años
hace 5 años
Hello people, I have a failure in rx siemens tridoros team, the issue is that when doing radioscopy in one hand it looks barbaric but when I do lumbar or another technique that I need more power you can not see anything, can you help me ?? Thank you
robertohoracio
hace 5 años

Hi. A query. The tests are in manual. What time has the intensifier. You use a grill of a serious photographer.

Responder
walservi40
hace 5 años

Hi Roberto, thank you very much for your help, if they are automatic, the IDI is about 30 years old, I think that since the equipment was installed, the grill is not being used. regards

Responder
robertohoracio
hace 5 años
hace 5 años
All right. If the idi is very old and the high source is fine. It may have poor performance. If the thickness to cross is large generates dispersed radiation you must use grill. This improves the image but increases the technique by 10%. Check the source of high idi but I think it is depletion of idi. Comment and greetings
walservi40
hace 5 años

Ok thank you very much Roberto, any news let you know !! regards

Responder

¿Conoces una solución?

¡Aprende a reparar con estos cursos YoReparo!

4.84
(25)

Curso Práctico de Reparación y Bobinado de Motores Eléctricos

Héctor Castiñeira
$ 60 USD
$ 20 USD
4.73
(22)

Curso Reparación de Fuentes Conmutadas de PC y Equipos Electrónicos

Maximiliano Resch
$ 40 USD
$ 20 USD
4.6
(10)

Curso Localización de Fallas en Refrigeración en Heladeras/Neveras On/Off

Yamil Alarcón
$ 29 USD
$ 12 USD
¿Tienes una mejor respuesta a este tema? ¿Quieres hacerle una pregunta a nuestra comunidad y sus expertos?
Regístrate gratis