hace 5 años
hace 5 años
Hello colleagues, I am new to electromedicine. although so far I have to solve electronic problems, but now I have a consultation with the cable of the dental x-ray tube head. My question is if it is a special cable? Or is it a common one. the fault that precenta is that when moving the head enters short. apparently he has the cable cut somewhere. what cable could i use ??? Thanks in advance
martin villegas
hace 5 años
hace 5 años
matias: usually the cables they use have serigraphic letters on the lining if you mark other letters more than tw or thw it can be special I recommend that you take pictures of how it is connected and remove it and checks where they sell wiring to see if It is special, another option is that you post a photo where you can read the letters that I mentioned to give you an idea of what type is handling greetings and comments
matias1232017
hace 5 años

Good your information. I have to take it apart next week. I'm going to take pictures so we can analyze it, thank you friend

Responder
martin villegas
hace 5 años

we are pending

Responder
matias1232017
hace 5 años

hello mate, it was the nomas cable, for a moment I thought it didn't fail before the shot, but then I realized that it has an open circuit alarm. thanks for the contributions

Responder
martin villegas
hace 5 años

greetings and I'm glad I supported you

Responder
junior259
junior259
103
hace 5 años
hace 5 años
Tube head cable ?, it would be good to look for a diagram and analyze the assembly of the head, if it measures you short in the terminals of the head, if there are two points you have to measure, it could be normal, since at that point you will find a very low ohmiaje, if by saying "short" you mean that the head is disconnected, it is most likely that some of the cables are indeed broken in some section of its extension ... If you are going to change the wiring, you can consider one that has good conductivity, is relatively flexible and, if possible, has good insulating thickness, since the movement of the arm and head is where the cables tend to break and consequently the equipment breakdown. Luck
matias1232017
hace 5 años

Today I will see it, thanks for the contribution. the same now the question arises that the cable is. Talk to the owner and say he blocks when he accommodates. If it were short on the cable, it would have to fail at the time of firing.

Responder
junior259
junior259
103
hace 5 años

If it is blocked by simply accommodating it, it depends on the equipment you are seeing, it can be blocked by detecting a short or disconnection of the head ... Most likely, the problem is in the cable and / or connection of the head, in equipment relatively old so that the head is not, or the cables are short, the equipment will not detect it, but in relatively modern equipment it may be that if it detects any short or disconnection in the head ...

Responder
junior259
junior259
103
hace 5 años

I would appreciate my comment if it helped you ... Consider that the ratings you give to each response or comment help us

Responder
Reparaciones similares solucionadas

¿Conoces una solución?

¡Aprende a reparar con estos cursos YoReparo!

4.25
(8)

Reparación de Placa Main TV LED con el Ing. Picerno

Ing. Alberto H. Picerno
$ 40 USD
$ 20 USD
4.94
(16)

Curso Refrigeración Comercial Aplicado a Diagnóstico y Reparación

Ing. Georga Melgarejo
$ 40 USD
$ 13 USD
4.64
(45)

Curso Tecnología Inverter - Fabricación de probadores y herramientas

Yamil Alarcón
$ 240 USD
$ 40 USD
¿Tienes una mejor respuesta a este tema? ¿Quieres hacerle una pregunta a nuestra comunidad y sus expertos?
Regístrate gratis