dental work ray equipment. fail, intermittent beep.

matias1232017
hace 5 años
hace 5 años
Hello colleagues I got a work ray team which precedes a failure when the head moves. It is an intermittent beep, there to turn it off so that the fault comes out. finally check the plate which is fine. disconnect the head to see if it was the problem, which now arises the doubt, to disconnect the head throws that fault. when I connect it, not when we move the head to all sides, which I said was the cable, but what I can't understand is that the fault appears with the head disconnected. if the head does not energize until the moment of firing. It's a good idea I listen to opinions, greetings
walservi40
hace 5 años

Hi Matias, it may be a false contact or you should set the level of oil that the header may have, the beep sounds for safety due to lack of oil!
Check it out and comment
regards
walterserviciotecnico @ gmail

Responder
junior259
junior259
103
hace 5 años
hace 5 años
Good morning estimates, at your personal discretion, if you only have two cables that come out of the head to be energized, the most likely problem is the connection, the system not detecting the head coil gives you the beeps ... So even if the wiring is broken at some point, in some position the connection opens and gives you the alarm ... The solution would only be to change the wiring, if the head rotation does not have a stop, the most likely is that the cable will become extrangle and fracture at some point ... luck ...
matias1232017
hace 5 años

hello, thank you for the opinions, well now I corroborate that if that fault exists then. the open circuit one can say, I was not sure that it existed, because in other equipment I saw there was no such fault, but hey, let alone that all the equipment is different.

Responder
junior259
junior259
103
hace 5 años

Score ... xD

Responder
Reparaciones similares solucionadas

¿Conoces una solución?

¡Aprende a reparar con estos cursos YoReparo!

4.25
(8)

Reparación de Placa Main TV LED con el Ing. Picerno

Ing. Alberto H. Picerno
$ 40 USD
$ 20 USD
4.94
(16)

Curso Refrigeración Comercial Aplicado a Diagnóstico y Reparación

Ing. Georga Melgarejo
$ 40 USD
$ 13 USD
4.64
(45)

Curso Tecnología Inverter - Fabricación de probadores y herramientas

Yamil Alarcón
$ 240 USD
$ 40 USD
¿Tienes una mejor respuesta a este tema? ¿Quieres hacerle una pregunta a nuestra comunidad y sus expertos?
Regístrate gratis