⚡ Descuentos en cursos en video - Lleva tu aprendizaje técnico al siguiente nivel y aprovecha precios especiales antes de que termine la oferta. 🔥 Aprovechar oferta ahora 🔥
Tomographs

Rx tube tomograph

hernansam
hace 5 años
hace 5 años
Hello how are you ?, I ask you a question I have a ct sytec 3000 that 2 years ago it is standing, without use, the equipment tube as it was so long without use, may I have a problem wanting to use it back?
gilrubio
gilrubio
583
hace 5 años
hace 5 años
It is possible that gas has formed inside the tube, what is recommended in these cases is a procedure called "seasoning", in the most modern equipment, they can already do it with a simple eating, in the case of your equipment it has to be done manually: you have to start making short shots, with low techniques with both filaments, 4 or 5 shots with each technique. Gradually see the technique going up, first the 40, 60, 80, 100, 150, 200 with each filament then raise the KV 100 and start again with low mA.