hace 5 años
What bothers me is its operation will be that it weighs well or badly, when I put a standard weight (KG or grs) it does not do what the other electronic scales do to climb from 1 in one rises from 10gr to 10gr
for example I have something of 145 gr and it sends it to 150 gr, and if I add something like for example 135 it rounds it to 140
- The first problem I had was that it was not zero, it was always changing, I changed a 1.0uf capacitor and bye problem
- replace the voltage regulators + and - with new ones. (just in case)
- adjust the internal calibration preset to zero.
I measured the cell in ohms and millivolts scale (2mv 350 ohms cell) (red, green, black and white wire)
TECHNICAL SPECIFICATIONS:
KRETZ 5580 30KG
Du = dp = $ 0.01
e = dd = dt = 10 g
Max = 30 g
Min = 0.2 kg
Tare = -15 kg
Limit = 30 kg
220v 50HZ
Argentina origin class III
I appreciate your help