Good colleagues He told them that this balance entered my workshop with the problem that was marked in the count (9) marked all 9 after checking, change the short communication cable between the monitor and the cell and started. But to my surprise, the balance does not weigh, it is always marking 0 and it turns on the stable of 0 and the function les (that is not normal for me to be on) but does not mark anything when a weight is placed on it. Or will it have to be recalibrated? In that case, I would like if you can give a hand in advance. Thank you very much.
enrique fernandez
hace 5 años
hace 5 años
check that the cell is good and that the cables are well connected, in the forum there is enough information about this balance, check this link http://yoreparo-prod.us-east-1.elasticbeanstalk.com/app.php/balanzas/preguntas/5814436/hola-necesito-manual-de-calibracion-de-balanza-xk-3190-a12
mac 2
mac 2
40
hace 5 años

Thank you colleague the link was helpful. The problem was that for some reason all the load parameters were deleted. I recalibrated it and it was perfect

Responder
Reparaciones similares solucionadas

¿Conoces una solución?

¡Aprende a reparar con estos cursos YoReparo!

4.73
(22)

Curso Reparación de Fuentes Conmutadas de PC y Equipos Electrónicos

Maximiliano Resch
$ 40 USD
$ 20 USD
4.71
(7)

Curso TV OLED y QLED con el Ing. Picerno

Ing. Alberto H. Picerno
$ 29 USD
$ 10 USD
4.64
(11)

Curso Electrónica para Reparación de Placas

Santiago Cediel
$ 29 USD
$ 11 USD
¿Tienes una mejor respuesta a este tema? ¿Quieres hacerle una pregunta a nuestra comunidad y sus expertos?
Regístrate gratis